The m/z ratio is an ion's mass divided by its charge, and it is the value a mass spectrometer actually measures. A peptide of mass M carrying z added protons appears at (M + z times 1.00728) divided by z. The same peptide can show several peaks, one for each charge state.
- m/z is an ion's mass divided by its number of charges, the quantity a mass spectrometer measures.
- For a peptide of mass M with z added protons, m/z = (M + z × 1.007276) / z.
- NIST's CODATA proton mass is 1.007 276 466 5789 u.
- Leucine enkephalin (monoisotopic 555.2693) is expected at 556.2766 as [M+H]+ and 278.6419 as [M+2H]2+.
- Neighboring peaks in an electrospray series differ by one charge, which lets you solve for z.
What does m/z mean?
A mass spectrometer never weighs a molecule directly. It pushes charged molecules through electric or magnetic fields and watches how they move, and how an ion moves depends on two things at once: how heavy it is and how much charge it carries. The number that comes out is mass divided by charge, written m/z.
Picture two carts rolling down a ramp that is pushed by a fan. A heavy cart with two sails moves like a lighter cart with one sail. The instrument sees only the ratio of weight to sail, never the two separately. A peptide ion with twice the charge lands at roughly half the m/z.
The m stands for mass in daltons (unified atomic mass units) and the z for the number of elementary charges on the ion. For peptides analyzed in positive mode, that charge almost always comes from extra protons picked up during ionization, so z is simply the number of protons added.
This page is the arithmetic behind a peptide identity check. For the method as a whole, see our guide to LC-MS peptide identity testing.
- Find the monoisotopic mass M
- Pick a charge state z
- Add z proton masses to M
- Divide the total by z
- Match the peak within mass accuracy
Why do peptides carry more than one charge?
Electrospray ionization is the usual way peptides are turned into gas-phase ions for LC-MS. In their 1989 paper in Science, Fenn and colleagues described what sets electrospray spectra of large molecules apart: coherent sequences of peaks whose ions are multiply charged, with each peak differing by one charge from its neighbors. They reported spectra for proteins with molecular weights up to 130,000.
That multiple charging is the reason electrospray works for big molecules on instruments with a limited m/z range. A protein of tens of thousands of daltons carrying dozens of protons lands at an m/z value of one or two thousand, well inside what the analyzer can measure.
For a small peptide the picture is simpler. A pentapeptide might show mostly one or two charge states. A peptide of thirty or forty residues with several basic side chains can show three, four or more. Each charge state is the same molecule, so each peak must fit the same mass once the charge is accounted for.
The m/z formula
For a peptide of neutral monoisotopic mass M that has picked up z protons, the expected m/z is:
m/z = (M + z × 1.007276) / z
The 1.007276 is the mass of a proton, not of a hydrogen atom. NIST's CODATA value for the proton mass is 1.007 276 466 5789 u. A hydrogen atom also carries an electron, so using the hydrogen atom's mass instead would put every calculated value slightly off, and at high resolution that small error is visible.
The ions are written in a standard shorthand:
- [M+H]+ is the peptide plus one proton, charge 1.
- [M+2H]2+ is the peptide plus two protons, charge 2.
- [M+3H]3+ is the peptide plus three protons, charge 3.
The M in the formula has to be the right kind of mass. Use the monoisotopic mass when comparing against a high-resolution spectrum, because the tallest peak for a small peptide is the one made of the most common isotopes. Average molecular weight is for weighing powder, and it gives the wrong answer here. Our molecular formula and weight guide explains the difference, and the amino acid residue page shows how M is built from residue masses.
Worked example: leucine enkephalin
Leucine enkephalin, Tyr-Gly-Gly-Phe-Leu, is a short, well-documented peptide that makes a good exercise. PubChem lists its formula as C28H37N5O7 and its monoisotopic mass as 555.26930. Using the proton mass of 1.007276:
| Ion | Charge z | Calculation | Expected m/z |
|---|---|---|---|
| [M+H]+ | 1 | (555.26930 + 1.00728) / 1 | 556.2766 |
| [M+2H]2+ | 2 | (555.26930 + 2.01455) / 2 | 278.6419 |
| [M+3H]3+ | 3 | (555.26930 + 3.02183) / 3 | 186.0970 |
Notice that the doubly charged ion is not at exactly half of 556.2766. Half of that would be 278.1383. The real value is higher because the second proton adds its own mass before the division. Halving the singly charged value is a common shortcut that puts the prediction about half a unit off.
Whether a five-residue peptide actually shows a 3+ ion depends on how many sites can hold a proton, and the table does not claim it will. It shows where each ion would appear if it forms. In a real spectrum you look for the predicted peaks and check that the observed values sit within the instrument's stated mass accuracy.
Going backward is just as easy. To recover M from an observed ion, multiply the m/z by z and subtract z protons. For the 3+ value, 186.0970 times 3 is 558.2911, and taking off 3.0218 returns 555.2693.
Working out the charge from two peaks
A spectrum does not label each peak with its charge, so you often have to deduce it. Fenn's observation gives the handle: in an electrospray series, neighboring peaks differ by exactly one charge. If a higher-m/z peak p1 carries charge n minus 1 and the next peak down, p2, carries charge n, then:
n = (p1 − 1.007276) / (p1 − p2)
Try it with the example. Take p1 = 278.6419 and p2 = 186.0970. The top line is 277.6346, the bottom line is 92.5449, and the ratio comes out at 3.000. So the lower peak is the 3+ ion and the higher one is 2+, which matches the table. Once n is known, M follows from either peak: 278.6419 times 2 is 557.2838, and removing two protons gives 555.2693, the same mass the 3+ ion returned. Two independent peaks landing on one mass is much stronger evidence than a single match.
Software does this for whole series at once, a step called deconvolution, and reports a single neutral mass. For large molecules it is the only practical way to read the spectrum. Doing it by hand for two peaks is still a useful check that the software's answer makes sense. It also shows you which peaks the software grouped together, which matters when two compounds of similar mass sit in the same run and their charge series overlap on the m/z scale.
The formula works in both directions along a series. With n settled, the next peak up the m/z scale must carry n minus 2 charges, and its predicted position can be calculated before you look for it. A peak that fits none of the predicted positions belongs to something else in the sample, such as a related impurity or a different compound, and it should be explained before the result is accepted.
Reading m/z on a certificate
A certificate of analysis that reports mass spectrometry usually gives an observed value beside a theoretical one. Before comparing them, check three things. First, whether the values are m/z for a named ion or a deconvoluted neutral mass. Second, whether the theoretical value is monoisotopic or average. Third, whether the charge state is stated. A reported 278.64 means little until you know it is the 2+ ion.
Check the arithmetic yourself when you can. Take the stated sequence, build M from the residue masses plus one water or copy it from a reference record, apply the formula for the stated charge, and compare. If your number and the certificate's theoretical value disagree, one of the inputs differs, often the choice of monoisotopic against average mass.
Mass confirms identity, not purity. A correct m/z shows the expected peptide is present, while HPLC purity tells you how much of the sample it makes up. Results for our own lots appear only once the certificate for that lot is published on the certificate index.
FOR LABORATORY AND IN-VITRO RESEARCH USE ONLY. NOT FOR HUMAN OR ANIMAL CONSUMPTION.
Common questions
What is the formula for m/z of a peptide?
Why does one peptide give several peaks?
Is [M+2H]2+ just half of [M+H]+?
Should I use monoisotopic or average mass for m/z?
More laboratory guides
Sources
- NIST CODATA Fundamental Physical Constants: proton mass in u. Proton mass 1.007 276 466 5789 u used in every m/z calculation.
- PubChem compound record: Leucine enkephalin. Formula C28H37N5O7 and monoisotopic mass 555.2693 used in the worked example.
- Fenn JB et al., Electrospray ionization for mass spectrometry of large biomolecules, Science 1989 (Europe PMC record). Multiply charged ion series in which neighboring peaks differ by one charge; proteins up to 130,000 in molecular weight.
- IUPAC-IUB JCBN, Nomenclature and Symbolism for Amino Acids and Peptides, 3AA-11 to 3AA-13. Definition of a peptide and peptide bond, amino-acid residues, N- and C-terminal residues, and how peptides are named.